• LS-100
  • LS-100

LS-100

No.LS-100
LS-100
GET A QUOTE
  • LS-100
  • 产品详情

  • Contact Us

The LS-100 grating spectroradiometer is specially designed for use in measuring the spectral irradiance of solar simulators that are used for the measurement of photovoltaic cells.

The LS-100 is calibrated against a NIST-traceable standard lamp. Its size is compact and it is very easy to operate.

The LS-100 must be connected to a PC (or laptop computer) via a RS232C cable, and data are obtained using the furnished (included) software. The measurement/analysis software is used to process the obtained data as follows;

 compare with the data obtained in the standard test conditions
 calculate the integrated irradiance in the specified spectral range
 calculate the spectral coincidence (JIS C 8912)
 calculate the difference and ratio compared to the standard data

Calibration Flow Chart
for LS-100 of Spectroradiometer
Standard Lamp
In conformity with NIST
Indoor comparison Artificial Light
Spectroradiometer
 
OL752 Solar Simulator
Reference Instrument LS-100
 
 
Subject (to be delivered)
Instrument LS-100
 

Specificastion LS-100
Spectral range

350 to 1050 nm

Spectral interval 3.2nm
Spectral resolution 10nm
Spectral accuracy < 0.3nm
Aperture angle (Full view angle) 30°
Stray light 0.15%
Exposure time 0.01 to 5 sec
Communication RS232C
Software Applicable to MS-Windows 95/98 SPDac-for measurement, SPMan-for data analysis Automatic calculation of spectral match accuracy (IEC904-9)
Unit of output W/m²µm
Power supply DC12V, 0.8A
Size / Weight W50 x D250 x H150 mm / 1.8kg
Shanghai Sales + Technical Support + After-sales Service Center
Rayscience Optoelectronic Innovation Co., Ltd

address: 4th Floor, Building 122, Lane 2338, No. 1 Duhui Road, Minhang District, Shanghai
Telephone:
021-34635258 021-34635259
fax:
021-34635260
E-mail:
saleschina@rayscience.com